Reduction of the patient’s received radiation dose to as low as reasonably achievable (ALARA) is based on
recommendations of radiation protection organizations such as the International Commission on Radiological
Protection (ICRP) and the National Radiological Protection Board (NRPB). The aim of this study was to explore the
frequency and characteristics of rejected / repeated radiographic films in governmental and private centers in Jenin city.
The radiological centers were chosen based on their high volume of radiographic studies. The evaluation was carried
out over a period of four months. The collected data were compiled at the end of each week and entered into a computer
for analysis at the end of study. Overall 5000 films (images) were performed in four months, The average repeat rate of
radiographic images was 10% (500 films). Repetition rate was the same for both thoracic and abdominal images (42%).
The main reason for repeating imaging was inadequate imaging quality (58.2%) and poor film processing (38%).
Human error was the most likely reason necessitating the repetition of the radiographs (48 %). Infant and children
groups comprised 85% of the patient population that required repetition of the radiographic studies. In conclusion, we
have a higher repetition rate of imaging studies compared to the international standards (10% vs. 4-6%, respectively).
This is especially noticeable in infants and children, and mainly attributed to human error in obtaining and processing
images. This is an important issue that needs to be addressed on a national level due to the ill effects associated with
excessive exposure to radiation especially in children, and to reduce cost of the care delivered.
Authors
Abed Al Nasser ASSI
Pages From
33
Pages To
36
Journal Name
Polish Journal of Medical Physics and Engineering
Volume
24
Issue
1
Keywords
patient dose; radiography; exposure; repeat rate.